Report an error   


Submitted: 16 May 2007 Modified: 12 May 2017
HERDIN Record #: PCHRDPC040672

Comparison of hemostasis, vessel bursting and pressures and wall damage of an alternative cautery shears, bi-polar electrosurgery and harmonic scalpel in porcine vessel model: A randomized study.

Deogracias Alberto G. Reyes,
Eziquel Yumang,
Edgardo B. Cabrera

See More

Objective: The efficacy of an alternative thermocautery shears (ATS) constructed by the investigators was assessed in a porcine model against an electrosurgical unit (ESU) and Harmonic scalpel (HS) Methods: Forty-five mesenteric arteries were divided using the respective apparatus. Time from application to transection and hemostasis was recorded. Thirty 1 centimeter segments of femoral artery were sealed by the test apparatus and bursting pressures obtained. The welded mesenteric vessels were evaluated for wall damage Results: No statistical difference between the HS and the ATS for time of vessel transection and hemostasis was noted. The bipolar ESU has the longest time (p0.001). Bursting pressures were comparable between the HS and the ATS (p Conclusion: This study showed that the ATS may be a viable alternative to current modalities of hemostasis.

Publication Type
Journal
Publication Sub Type
Journal Article, Reprint
Title
Philippine Journal of Surgical Specialties
Frequency
Quarterly
Publication Date
July 2004
Volume
59
Issue
3
Page(s)
108-114
LocationLocation CodeAvailable FormatAvailability
Philippine Council for Health Research and Development Library Box No 6 Fulltext pdf (Request Document)
1. Reyes , DAG , Tabira, J , Galang, C . "Comparison of an alternative electrocautery apparatus with a commercially available electrocautery unit in terms of immediate coagulative ability and extent of tissue in jury" Philipi J Surg Spec 58((1)): 27-31 , 2003
2. 2. Biber , EJ , Redwine , DB , Tucker, RD et al. "An in-vivo evaluation of the correlation between histologic and visually estimated zones of tissue injury created with monolpolar elextorsurgery" J AM Assoc Gynecol Laparosc 1((4)Part2): S3-4, 1994. Retrieved from: http://www.ncbi.nlm.nih.gov/pubmed?term=An%20in-vivo%20evaluation%20of%20the%20correlation%20between%20histologic%20and%20visually%20estimated%20zones%20of%20tissue%20injury%20created%20with%20monolpolar%20elextorsurgery
3. Lantis, JC , Durville, FM et al. "Comparison of coagulation modalities in surgery" J Lapareondosc Adj Surg Tach A 8((6)): 381-394, 1998
4. Landman, J , Kerbi, K , Rehman, J et al. "Evaluation of a vessel sealing system,bipolar electrosurgery, harmonic scalpel,titanum clips, endoscopic gasttrointestinal anastomosis , vascular staples and sutures for arterial and vessel ligation in porcine model" Division of Urology, Department of Surgery Washington University School of Medicine St. Louis, Missouri USA , . Retrieved from: http://www.ncbi.nlm.nih.gov/pubmed?term=Evaluation%20of%20a%20vessel%20sealing%20system%2Cbipolar%20electrosurgery%2C%20harmonic%20scalpel%2Ctitanum%20clips%2C%20endoscopic%20gasttrointestinal%20anastomosis%20%2C%20vascular%20staples%20and%20sutures%20for
5. Michaletz, PA , Judge, D . "Microwave energy compared with heater probe and BICAP in an canine models of peptic ulcer hemorrhage " Gastroenterology 97((3)): 676-84, 1989. Retrieved from: http://www.ncbi.nlm.nih.gov/pubmed?term=Microwave%20energy%20compared%20with%20heater%20probe%20and%20BICAP%20in%20an%20canine%20models%20of%20peptic%20ulcer%20hemorrhage
6. Treat, MR , Bass, LS . "Tissue welding: a comprehensive clinical review" Lasers in surgery and Medicine , 1995. Retrieved from: http://www.ncbi.nlm.nih.gov/pubmed?term=Tissue%20welding%3A%20a%20comprehensive%20clinical%20review
7. Abstein, GT . "Practical electrosurgery for clinicians" Professional Medical EducationAssociation Inc. , 1999
8. Amaral, JF , Chrostek, CA . "Experimental comparison of the ultrasonically activated scalpel and laser surgery for laparoscopic use" Minimally Invasive Therapeutics and Allied Technology 6, 324-331, 1997
9. Deo, SV , Shukla, NK , Asthana, S , Niranjan, B , Srinivas, G . "A Comapative study of modified radical mastectomy using harmonic scalpela and electrocautery. " Department of Surgical Oncology. Institute Rotary Cancer Hospital, All India Institure of Medical Sciences, New Delhi ,
10. Armstrong, DN , Ambrose, WL , Schertzer, ME , Orangio, GR . "Harmonic scalpel vs. electrocautery hemorroidectomy: a prospective evaluation " Georgia Colon and Rectal Clinic Atlanta 30342, USA , . Retrieved from: http://www.ncbi.nlm.nih.gov/pubmed?term=%22Harmonic%20scalpel%20vs.%20electrocautery%20hemorroidectomy%3A%20a%20prospective%20evaluation%20